Horizontal Pulse Vacuum Autoclave, Sliding Door

  • Model : LST-PBH120PV; LST-PBH150PV
  • Chamber Design Pressure : -0.1/0.3MPa
  • Jecket Design Pressure : 0.3MPa
  • Generator Design Pressure : 1.0MPa
  • Design Temperaure : 150℃
Category:

Description

Horizontal Pulse Vacuum Autoclave Feature:

  1. Motor-driven sliding door for automatic opening and closing, no manual operation.
  2. Double-door design with pressure and temperature safety interlocks.
  3. Inflatable pressure-driven silicone seal, wear-resistant with up to 1500 cycles lifespan.
  4. Rectangular structure with circular reinforcement for strong sealing.
  5. Flash steam vacuum drying with fabric drying rate <0.6%, instrument drying <0.1%.
  6. Real-time process data printing, optional paperless recorders for pharmaceutical use.

Horizontal Pulse Vacuum Autoclave Specifications:

Model LST-PBH120PV LST-PBH150PV
Chamber Size (ØxL)mm 680×1180×1500 680×1180×1800
Volume (L) 120 150
Overall Sizes (WxDxH)(mm) 2007×2110×2113 2357×2110×2113
Weight (kg) 2300 2450
Steam Consumption (KG/C) 600 750
Water Consumption (KG) 65 80
Power 380V, 50HZ, 3.5KW(72KW) 380V, 50HZ, 4KW(90KW)
N.W.(kg) 2300 2450
Chamber Design Pressure -0.1/0.3MPa
Jecket Design Pressure 0.3MPa
Generator Design Pressure 1.0MPa
Design Temperaure 150℃
Maximum Operating Temperature 139℃
Maximum Operating Pressure 0.25MPa
Chamber Relief Valve 0.28MPa
Jacket Relief Valve 0.28MPa
Accuracy of Temperature 0.1℃
Accuracy of Pressure 0.1KPa
Sterilization Temperature Control Accuracy ±1℃
Temperature Uniformity ±1℃
Negative Pressure Pulsating Vacuum -80~0KPa
Across the Pressure Pulsation -80~80KPa
Positive Pressure Pulse 0~9 times
Maximum Vacuum -96KPa