- Model : LST-PBA650PV; LST-PBA800PV; LST-PBA1000PV; LST-PBA1200PV; LST-PBA1500PV;
- Volume(L) : 650 / 800 / 1000 / 1200 / 1500
- Chamber Design Pressure : 0.3MPa
- Generator Design Pressure : 1.0MPa
- Design Temperaure : 150℃
Horizontal Pulse Vacuum Autoclave, Motor Hinge Door
Description
Horizontal Pulse Vacuum Autoclave Feature:
- Motorized chain-driven sliding door lifts up and down.
- Double door interlock, including liquid pressure temperature safety interlock.
- Inflatable sealing ring, wear-resistant silicone, life 1500 times.
- Rectangular main structure, no flange European style reinforcement design.
- Siemens PLC control, touch screen display full automation.
- IOT remote monitoring and maintenance, support real-time data, troubleshooting, remote upgrade.
- Negative pressure, cross and positive pressure pulsation, air exclusion rate >99.99%.
- Flash vacuum drying, fabric drying rate <0.6%, apparatus <0.1%.
- Built-in printer prints data in real time, suitable for pharmaceutical customers.
- Sanitary stainless steel piping, using German-made control valve and water ring vacuum pump.
- Supply system detection alarm, automatically prevent program startup failure.
- Automatic drainage temperature control, drainage temperature <60℃.
Horizontal Pulse Vacuum Autoclave Specifications:
Model | LST-PBA650PV | LST-PBA800PV | LST-PBA1000PV | LST-PBA1200PV | LST-PBA1500PV |
Chamber Size (ØxL) (mm) | 610×910×1200 | 610×910×1500 | 610×910×1800 | 680×1180×1500 | 680×1180×1850 |
Volume(L) | 650 | 800 | 1000 | 1200 | 1500 |
Overall Sizes (WxDxH)(mm) | 1500×1350×1900 | 1800×1350×1900 | 2100×1350×1900 | 1800×1420×2000 | 2100×1420×2000 |
Weight (kg) | 1250 | 1350 | 1550 | 1650 | 1850 |
Steam Consumption (KG/C) | 35 | 47 | 55 | 65 | 80 |
Water Consumption (KG) | 320 | 400 | 500 | 600 | 750 |
Power | 380V, 50Hz 3KW+(36KW) |
380V, 50Hz 3.5KW+(48KW) | 380V, 50Hz 3.5KW+(54KW) | 380V, 50Hz 3.5KW+(72KW) | 380V, 50Hz 4KW+(90KW) |
N.W. (kg) | – | 1350 | 1550 | 1650 | 1850 |
Chamber Design Pressure | 0.3MPa | ||||
Generator Design Pressure | 1.0MPa | ||||
Design Temperaure | 150℃ | ||||
Maximum Operating Temperature | 139℃ | ||||
Maximum Operating Pressure | 0.25MPa | ||||
Chamber Relief Valve | 0.28MPa | ||||
Jacket Relief Valve | 0.28MPa | ||||
Accuracy of Pressure | 0.1KPa | ||||
Sterilization Temperature Control Accuracy | ±1℃ | ||||
Temperature Uniformity | ±1℃ | ||||
Negative Pressure Pulsating Vacuum | -80~0KPa | ||||
Across the Pressure Pulsation | -80~80KPa | ||||
Positive Pressure Pulse | 0~9 times | ||||
Maximum Vacuum | -96KPa |
WHY Labpeak
- 24 Months Limited Warranty
- CE & FDA Standards
- 1-2 Days Handling and 7-30 Days Shipping